Influence of lead apron shielding on absorbed doses from cone-beam computed tomography

Dennis Rottke, Jonas Andersson, Ken Ichiro Ejima, Kunihiko Sawada, Dirk Schulze

研究成果: ジャーナルへの寄稿記事査読

6 被引用数 (Scopus)

抄録

The aim of the present work was to investigate absorbed and to calculate effective doses (EDs) in cone-beam computed tom ography (CBCT). The study was conducted using examination protocols with and without lead apron shielding. A full-body male RANDO(r) phantom was loaded with 110 GR200A thermoluminescence dosemeter chips at 55 different sites and set up in two different CBCT systems (CS 9500(r), ProMax(r) 3D). Two different protocols were performed: The phantom was set up (1) with and (2) without a lead apron. No statistically significant differences in organ and absorbed doses from regions outside the primary beam could be found when comparing results from exposures with and without lead apron shielding. Consequently, calculating the ED showed no significant differences between the examination protocols with and without lead apron shielding. For the ProMax(r) 3D with shielding, the ED was 149 pSv, and for the examination protocol without shield ing 148 pSv (SD = 0.31 pSv). For the CS 9500(r), the ED was 88 and 86 pSv (SD = 0.95 pSv), respectively, with and without lead apron shielding. The results revealed no statistically significant differences in the absorbed doses between examination with and without lead apron shielding, especially in organs outside the primary beam.

本文言語英語
論文番号ncw275
ページ(範囲)110-117
ページ数8
ジャーナルRadiation Protection Dosimetry
175
1
DOI
出版ステータス出版済み - 1 6月 2017

フィンガープリント

「Influence of lead apron shielding on absorbed doses from cone-beam computed tomography」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル