Head-to-head comparison of 18F-sodium fluoride coronary PET imaging between a silicon photomultiplier with digital photon counting and conventional scanners

Hidenobu Hashimoto, Keiichiro Kuronuma, Mark C. Hyun, Donghee Han, Valerie Builoff, Sebastian Cadet, Damini Dey, Daniel S. Berman, Jacek Kwiecinski, Piotr J. Slomka

研究成果: ジャーナルへの寄稿記事査読

3 被引用数 (Scopus)

抄録

Background: We compared silicone photomultipliers with digital photon counting (SiPM) and photomultiplier tubes (PMT) positron emission tomography (PET) in imaging coronary plaque activity with 18F-sodium fluoride (18F–NaF) and evaluated comprehensively SiPM PET reconstruction settings. Methods: In 25 cardiovascular disease patients (mean age 67 ± 12 years), we conducted 18F–NaF PET on a SiPM (Biograph Vision) and conventional PET (Discovery 710) on the same day as part of a prospective clinical trial (NCT03689946). Following administration of 250 MBq of 18F–NaF, patients underwent a contrast-enhanced CT angiography and a 30-min PET acquisition in list-mode on each PET consecutively. Image noise was defined as mean standard deviation of blood pool activity within the left atria. Target-to-background ratio (TBR) and signal-to-noise ratio (SNR) were measured within the whole-vessel tubular three-dimensional volumes of interest on the cardiac motion and attenuation-corrected 18F–NaF PET images using dedicated software. Results: There were significant differences in image noise and background activity between the two PETs (Image noise (%), PMT: 7.6 ± 3.7 vs SiPM: 4.0 ± 2.3, P < 0.001; background activity, PMT: 1.4 ± 0.4 vs SiPM: 1.0 ± 0.3, P < 0.001). Similarly, the SNR and TBR were significantly higher in vessels scanned with the SiPM PET (SNR, PMT: 16.3 ± 11.5 vs SiPM: 32.7 ± 29.8, P < 0.001; TBR, PMT: 0.8 ± 0.4 vs SiPM: 1.1 ± 0.6, P < 0.001). SiPM PET image reconstruction with a 256 matrix, 1.4 mm pixel, and 2 mm Gaussian filter provided best trade off in terms of maximal SNR, TBR, and clinically practical file size. Conclusions: In 18F–NaF coronary PET imaging, the SiPM PET showed superior image contrast and less image noise compared with PMT PET.

本文言語英語
論文番号102045
ジャーナルJournal of Nuclear Cardiology
42
DOI
出版ステータス出版済み - 12月 2024

フィンガープリント

「Head-to-head comparison of 18F-sodium fluoride coronary PET imaging between a silicon photomultiplier with digital photon counting and conventional scanners」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル