Balloon pin-hole rupture during percutaneous coronary intervention for recurrent, calcified in-stent restenosis: A case report

Nobuhiro Murata, Tadateru Takayama, Takafumi Hiro, Atsushi Hirayama

研究成果: ジャーナルへの寄稿記事査読

9 被引用数 (Scopus)

抄録

Percutaneous coronary intervention (PCI) for patients with in-stent restenosis (ISR) is generally considered safe and effective. However, due to increased tissue hardness, PCI for calcified intra-stent ISR is technically challenging. Here, we report severe angioplasty-related complications in a patient presenting with calcified, recurrent ISR following PCI. After receiving drug-coated balloon (DCB) angioplasty for an initial ISR, the patient developed recurrent ISR during the follow-up period. Intravascular imaging revealed intra-stent calcifications and balloon angioplasty was subsequently performed. During the angioplasty, a pin-hole balloon rupture occurred, consequently causing coronary dissection as visualized by intravascular imaging. To prevent acute coronary occlusion, stent implantation was required. The present case report suggests that, following detection of intra-stent calcified stenosis, both careful balloon inflation as well as optimal ablation device selection are required to prevent potential complications and obtain successful procedural outcomes.

本文言語英語
ページ(範囲)1287-1290
ページ数4
ジャーナルCatheterization and Cardiovascular Interventions
91
7
DOI
出版ステータス出版済み - 1 6月 2018

フィンガープリント

「Balloon pin-hole rupture during percutaneous coronary intervention for recurrent, calcified in-stent restenosis: A case report」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル