N2 - Colonoscopies are widely available, but there are cases where insertion can be difficult, even for experienced endoscopists. EC-760XP/L, a new ultrasmall-diameter long scope, may be useful in such cases. This single-center retrospective study included 39 cases where colonoscope insertion was difficult even when previously conducted by an experienced endoscopist. The primary outcome was the cecal intubation time using EC-760XP/L compared to the time used in a previous examination with a standard scope. The secondary outcomes were the cecum intubation rate, intestinal cleanliness level, adenoma detection rate, polyp detection rate, sedative use rate, occurrence of adverse events, and pain experience. A comparison of cecal intubation times between EC-760XP/L and the standard scope showed that insertion times were significantly lower with EC-760XP/L (9.5 min) compared to the standard scope (19 min) (p < 0.01). The standard scope achieved cecal intubation in 30 cases (76.9%), whereas EC-760XP/L reached the cecum in all cases (p < 0.01). Pain was observed in 3 cases (8.3%) with the EC-760XP/L, which was significantly lower than the 22 cases (56.4%) with the standard scope (p < 0.01). In conclusion, EC-760XP/L proved to be useful in cases where colonoscope insertion was difficult.
AB - Colonoscopies are widely available, but there are cases where insertion can be difficult, even for experienced endoscopists. EC-760XP/L, a new ultrasmall-diameter long scope, may be useful in such cases. This single-center retrospective study included 39 cases where colonoscope insertion was difficult even when previously conducted by an experienced endoscopist. The primary outcome was the cecal intubation time using EC-760XP/L compared to the time used in a previous examination with a standard scope. The secondary outcomes were the cecum intubation rate, intestinal cleanliness level, adenoma detection rate, polyp detection rate, sedative use rate, occurrence of adverse events, and pain experience. A comparison of cecal intubation times between EC-760XP/L and the standard scope showed that insertion times were significantly lower with EC-760XP/L (9.5 min) compared to the standard scope (19 min) (p < 0.01). The standard scope achieved cecal intubation in 30 cases (76.9%), whereas EC-760XP/L reached the cecum in all cases (p < 0.01). Pain was observed in 3 cases (8.3%) with the EC-760XP/L, which was significantly lower than the 22 cases (56.4%) with the standard scope (p < 0.01). In conclusion, EC-760XP/L proved to be useful in cases where colonoscope insertion was difficult.
KW - Adenoma detection rate
KW - Blue laser imaging
KW - Colorectal cancer
KW - Linked color imaging
KW - Ultrathin endoscopes
UR - http://www.scopus.com/inward/record.url?scp=85203997309&partnerID=8YFLogxK
U2 - 10.1038/s41598-024-72689-1
DO - 10.1038/s41598-024-72689-1
M3 - Article
C2 - 39277678
AN - SCOPUS:85203997309
SN - 2045-2322
VL - 14
JO - Scientific Reports
JF - Scientific Reports
IS - 1
M1 - 21506
ER -