Two revision surgeries on cemented custom-made tumor prostheses

Yukihiro Yoshida, Yuki Okamura, Mamoru Akita, Masashi Taniguchi, Toshio Kojima, Eiji Osaka, Yasuaki Tokuhashi

Research output: Contribution to journalArticlepeer-review

Abstract

We performed revision surgery in 2 patients for stem fracture of a cemented tumor prosthesis that occurred more than 25 years after the initial surgery. For revision, the global modular replacement system (GMRS) was used. However, as bone cement in the bone could not be adequately removed, stems with respective diameters of 11 and 12.5 mm were used. In revision surgery for cemented tumor prostheses, adequate removal of residual bone cement is optimal. However, when there is a risk of fracture, it may be appropriate to insert a thicker stem after reaming the femoral canal as much as possible, and then fix the stem using the cement-in-cement method.

Original languageEnglish
Pages (from-to)69-72
Number of pages4
JournalActa Medica Okayama
Volume71
Issue number1
Publication statusPublished - 2017

Keywords

  • Custom-made tumor prosthesis
  • Malignant bone tumor
  • Revision

Fingerprint

Dive into the research topics of 'Two revision surgeries on cemented custom-made tumor prostheses'. Together they form a unique fingerprint.